External Interface

Inputs and Outputs

The UNI provides two configurable input/output (IO) channels that can be used for measuring voltages, counting pulses, measuring frequency and duty cycle, controlling external devices, and interfacing to 1-Wire sensors.

Pin

Name

Description

4

IO1

Digital input, pulse input, 1-Wire, and output

5

IO2

Voltage input, and output

The available functions are summarised below.

Function

IO1

IO2

Digital measurement

X

Voltage measurement

X

Pulse count

X

Frequency

X

Duty cycle

X

Wake on change

X

Switch to GND

X

X

Permanent pullup

X

1-Wire

X

General IO Settings

Each IO can be configured independently.

Measurement Interval

Measurements can be scheduled as a multiple of the base interval. The fastest possible measurement rate is achieved by setting the Interval to 1, in which case measurements will occur on every base interval. To reduce power consumption, the measurement rate can be reduced by increasing the Interval. To disable an IO, set the Interval to 0.

Default State

The Default State sets the output state of an IO between measurement cycles: OFF or GND, or 1WIRE on IO1.

Note

The state of an IO may deviate from the settings if being controlled by a script.

Measurement State

The Measurement State sets the state of an IO during a measurement cycle: No Change, OFF or GND.

Measurement Time

The Measurement Time sets how long the device holds the IO in the measurement state before taking a measurement. This allows attached sensors to boot, or for measurements to settle before being measured.

IO1

IO1 supports digital measurement, pulse counting, frequency measurement, duty cycle measurement, wake-on-change, 1-Wire communications, and output switching.

Digital inputs are measured using a threshold of approximately 2 V. Voltages below the threshold are interpreted as OFF and voltages above the threshold are interpreted as ON.

Pulse counting can be used with flow meters, rain gauges, energy meters, run sensors, and other pulse-output devices. Frequency and duty cycle measurements use the same pulse measurement circuitry.

IO1 has a permanent 2.67k ohm pullup resistor to 3.3 V for 1-Wire signalling and digital input operation. This pullup is always connected, including when IO1 is configured as an input or set to OFF, and cannot be disabled in the settings or from a script.

Many pulse-output sensors, including reed-switch flow meters and rain gauges, use open-collector outputs or voltage-free contacts that switch to ground. The permanent pullup provides the bias required for pulse detection, so these sensors can be connected without enabling an additional pullup.

Note

A load connected between IO1 and GND will carry approximately 1 mA from the permanent pullup when IO1 is set to OFF. This may cause LED indicators to glow faintly. Loads switched by IO1 should be connected between a supply and IO1, and switched using the GND state.

For low-power applications, turn on Wake on Change for IO1 to wake the device when the input changes state.

IO2

IO2 supports voltage measurement and output switching.

Voltage measurements may be calibrated into end-use engineering units. For example, a 0-10 V level transmitter representing a water level of 0-1 m can be configured so that the measured value is reported directly in metres. This allows sensor outputs to be displayed and alarmed in meaningful application units rather than raw voltages.

Output Functions

The IO can also be used as outputs and can be switched to:

  • OFF - Disables output switching. IO1 retains its permanent pullup; IO2 is high impedance.

  • GND - Connects the IO to ground.

OFF disables the switched output functions. IO2 is left in a high impedance state, but IO1 remains biased by its permanent 2.67k ohm pullup to 3.3 V. An unconnected IO1 will therefore normally read ON. Connected sensors must be able to pull IO1 below the digital input threshold to produce an OFF reading.

GND connects the IO to ground through a protected low-side switch and is typically used to activate relays, alarms, indicators, contactors, and other externally powered devices.

Output Characteristics

Each IO output can switch voltages up to 32 V and sink currents up to 250 mA.

The outputs are protected against overloads, short circuits, electrostatic discharge, inductive transients, and excessive temperature. If a fault condition occurs, the output automatically protects itself and will automatically resume operation when the fault is removed.

Example of switching an resistive load

Switching a resistive load

The IO are capable of switching inductive loads such as relays and solenoids; however, flyback diodes should always be fitted to inductive loads to minimise electrical stress and improve system reliability.

Example of switching an inductive load

Switching a high inductance load

Pulse Counting

IO1 can be used to measure pulse count, frequency, and duty cycle.

Pulse counting records the number of input transitions occurring during the measurement period. Frequency measurement determines the pulse repetition rate, while duty cycle measurement determines the proportion of time the signal remains active.

Input transitions are detected when the signal crosses the digital input threshold of approximately 2 V.

Note

Duty cycle is measured at the approximately 2 V digital input threshold. For signals with significant rise and fall times, the measured duty cycle will depend on where the signal crosses this threshold.

Diagram showing how rise time affects duty cycle measurement

Effect of Rise and Fall Time on Duty Cycle Measurement

For reliable operation, input signals should transition cleanly through the threshold voltage and excessive noise around the threshold should be avoided.

The figure shows how to wire a flow meter with a reed-switch using the internal pullup on the UNI-Z1.

Diagram showing connections to measure pulses

Pulse counting

Wake on Change

Wake-on-change allows the UNI to wake from low-power sleep modes when IO1 changes state. Turn on Wake on Change for IO1 to use it.

This feature is particularly useful when monitoring alarm contacts, limit switches, float switches, reed switches, pulse-output sensors, and other event-driven inputs that require immediate response. The permanent pullup resistor on IO1 allows a connected contact or open-collector sensor to pull the input low when activated, without enabling an additional pullup.

1-Wire

IO1 can be configured for 1-Wire communication by selecting 1WIRE as the IO1 Default State. When 1-Wire is enabled, IO1 is dedicated to 1-Wire operation and all other IO1 measurement and output functions are disabled.

1-Wire is commonly used by digital temperature sensors, identification devices, and other low-speed digital sensors.

The 1-Wire data line is connected to IO1. The permanent 2.67k ohm pullup to 3.3 V provides the bus bias for 1-Wire signalling. Multiple 1-Wire devices can share the same bus as each device contains a unique factory-programmed address.

1-Wire device wiring

iButton reader connected to IO1

Communication with 1-Wire devices is performed entirely from script. The script is responsible for discovering devices, selecting addresses, issuing commands, reading responses, and dispatching measurements.

Note

Some 1-Wire devices support parasite power operation. Where reliable operation is important, powering sensors from an external supply is recommended.

Scripting

The IO can be controlled directly from scripts.

Scripts can read measurements and change output states.

Warning

If an IO is enabled in the settings and is also used in a script, the behaviour of that IO will be affected by both the settings and the script.

Note

In sleep and hibernate modes, the outputs default to OFF. The permanent 2.67k ohm pullup on IO1 remains connected.

Specification

Parameter

Specification

IO1 Characteristics

Open circuit voltage maximum

32 V

Pullup resistor (always on)

2.67k ohm to 3.3 V

IO voltage guaranteed to be detected as low

< 0.5 V

IO voltage guaranteed to be detected as high

>2.5 V

IO2 Characteristics

Open circuit voltage maximum

32 V

Resistance to ground

110k ohm

Output (IO1 and IO2)

GND sink current maximum

250 mA per pin - short-circuit protected, automatic recovery

Voltage Measurement (IO2)

Voltage ADC Type

12 bit SAR

Voltage range

0-32VDC

Voltage measurement resolution

10 mV

Voltage measurement accuracy

TBD

Threshold for IO to wake from hibernate

2.0 V +/-0.5V

Frequency and Pulse (IO1)

Frequency measurement range

1 Hz to 10 kHz

Frequency measurement accuracy

1 Hz

Pulse counting minimum frequency

0 Hz

Pulse counting maximum frequency - always on

10 kHz

Pulse counting maximum frequency - sleep

100 Hz *

Pulse counting pulse width minimum - always on

300usec

Pulse counting pulse width minimum - sleep

5msec

Duty cycle measurement range

2% to 98% (tested at 100 Hz)

Duty cycle accuracy

+-1% (tested at 100 Hz)

Note

The maximum pulse counting frequency in sleep assumes a 50% duty-cycle square wave. Higher pulse counting frequencies may be supported at the expense of increased power consumption. Contact support@senquip.com if higher pulse counting frequencies are required during sleep.

Serial interface

The serial port can be used to capture data that is sent from an external system or to interface to a MODBUS sensor.

The UNI-Z1 provides an RS485 interface on pins 2 and 7 of the header. The pin functions are shown in the table below. When required, an optional 120Ω termination resistor can be enabled in the settings.

Pin

Function

2

RS485-B

7

RS485-A

Note

RS485-B is sometimes referred to as D+ or TX+/RX+ and RS485-A as D- or TX-/RX-.

The serial interface supports RS485 operation.

The interface incorporates protection features to improve robustness and reliability in industrial environment with the RS485 receiver offering full failsafe operation for floating, shorted, or terminated inputs.

Specification

Parameter

Specification

RS485 differential output voltage (minimum with load resistance 120Ω)

+2 V

RS485 differential input threshold (typical)

+/-200mV

Maximum nodes in RS485 mode

256

RS485 termination resistor

120Ω

ESD protection

+/-16kV

Settings

Measurements can be scheduled as a multiple of the base-interval. The fastest possible measurement rate is achieved by setting the Interval to 1 in which case measurements will occur on every base interval. To reduce power consumption, the measurement rate can be turned down by increasing the Interval.

In serial capture mode the measurement interval can be used to reduce the number of readings being provided by a connected sensor or system that may be permanently powered. If for instance, a connected system is sending a message every second but it is only required to be read and transmitted every minute, the measurement interval can be set to 1 minute in which case the device will wake on the minute interval, receive a message and return to sleep thereby missing the other 59 messages sent by the attached system. Since serial packets cannot be interrogated by the UNI-Z1 without a customised script, it makes sense to set the measurement interval to the same as the transmit interval in most cases.

An optional 120Ω RS485 termination resistor can be selected by turning on Termination Resistor. The purpose of the termination resistor is to match the impedance of a transmission line to the hardware impedance of the interface to which it is connected. Termination is generally not required in lower speed networks (9600 baud or less) and networks shorter than 500 m in length. No more than 2 termination resistors should be used, one at each end of the RS485 transmission line.

A baud rate of 4800, 9600, 19200, 38400, 56800 or 115200 needs to be selected using the Baud Rate option. Other settings, including the number of bits, odd or even parity and 1 or 2 stop bits are added in the Settings field. The most common setup is 8 bits, no parity and 1 stop bit or “8N1”.

The serial interface can be configured in serial capture mode or MODBUS mode using the Mode option.

Set Mode to Capture, Modbus or Scripted. In Scripted mode the serial port is controlled entirely from a script.

Capture mode is typically used where an external sensor sends serial data and a portion of that serial data is to be captured. MODBUS mode is used to connect to external sensors that are compatible with the MODBUS standard.

In serial capture mode, the device listens for periodic data and when received, transmits this data at the next send interval. The maximum length of a message that can be captured is 512 characters. Once 512 characters have been received, the UNI-Z1 will terminate the capture and will transmit it on the next transmit interval.

In capture mode, the Max Time setting can be used to set a timeout after which the serial port will return to sleep. Max Time can be used as a way to end serial measurement in the event that no serial data is received, or as a mechanism to allow the device to sample the serial port for a defined time-period.

Note

If the serial port needs to be kept on all the time, set the Max Time to longer than the measurement interval. The contents of the serial buffer is retained as long as the device does not return to sleep.

The operation of the Max Chars option is similar to the Max Time setting except that the serial port stops sampling after a certain number of characters has been received. In most cases where the Max Chars setting is used to terminate serial capture, the Max Time setting is also used to end the serial measurement in the event that data does not arrive.

In serial capture mode, in systems where many messages are sent and only a few are of interest, a Start String of up to 10 characters can be enabled. For instance, in a typical GPS serial NMEA feed, the following are a subset of available messages:

  • DTM - Datum being used.

  • GGA - Fix information

  • GLL - Lat/Lon data

  • GSA - Overall Satellite data

  • GSV - Detailed Satellite data

  • RMC - Recommended minimum data for GPS

  • RTE - Route message

  • VTG - Vector track an Speed over the Ground

If in the application, the user is only interested in receiving the GGA message, then a Start String can be set to GGA. In that way, any messages starting with DTM, GLL, GSA or other unwanted messages will be discarded.

Note

If a start string is enabled, the device will stay awake until the string is received or until the Max Time is reached.

In firmware revisions less than 2, serial start strings are specified as text, with special characters such as carriage return and line feed being specified by their respective escape sequences. A list of allowable escape sequences is given below:

  • \f Form-feed

  • \n Newline (Line Feed)

  • \r Carriage Return

  • \t Horizontal Tab

  • \v Vertical Tab

  • \\ Backslash

Note

Because escape sequences start with a backslash (\), if a capture string contains a backslash, it needs to be escaped and so is represented as a double backslash (\\).

In firmware release 2 and above, serial start strings are specified as text, with special characters such as carriage return and line feed being specified by their respective ASCII codes in hexadecimal. A list of example hexadecimal sequences is given below:

  • \x0C Form-feed

  • \x0A Newline (Line Feed)

  • \x0D Carriage Return

  • \x09 Horizontal Tab

  • \x0B Vertical Tab

  • \x08 Backslash

The change to the method used to represent special characters has been made to allow for all ASCII characters to be used, and to allow for hexadecimal data to be captured.

Note

In firmware revisions 2 and lower, special characters are specified as escape characters. In revisions 2 and above, special characters are represented by their ASCII representations in hexadecimal.

In some serial protocols, the start of a packet is specified by a preceding period of inactivity on the serial bus. The Idle Time Before Start parameter can be used to specify an idle time, which is exceeded will trigger the serial port to start capturing serial data.

Note

If the serial port is capturing data and a subsequent idle time occurs, the capture process will restart and captured data will be discarded.

A serial capture End String of up to 10 characters can also be provided. Again using the NMEA example, all NMEA messages end with a carriage return and line feed and so the serial capture stop strings in each case will be the same and will be “\r\n” or \x0D\x0A in revision 2 and above firmware. In most instances, the serial stop strings will be the same for all messages.

Note

If a Start String is specified without a End String, or the End String is never encountered, the serial port will capture characters until the Max Time or Max Chars is reached, the next measurement interval occurs or 256 characters are received.

An optional serial Request String can be sent, on each measurement interval, to an external device. The purpose of the Request String is to request data from an external sensor or system. The Request String can be a maximum of 10 characters and can be entered as text. Special characters like carriage return and line feed can be inserted using escape sequences or their ASCII representations as described earlier in the chapter.

The UNI-Z1 implements the MODBUS communications protocol standard as a master, which enables communication with many slave devices connected to the network. The UNI-Z1 can be configured to periodically request specific data from slave MODBUS devices on the network and transmit that data at specified intervals.

Up to fifty MODBUS data requests can be configured; these can be from individual slave devices or several requests to the same device. Each request is a row in the Modbus table, with a Slave Address, Function and Register Address. The Slave Address is set by the manufacturer of the connected device; some devices allow it to be changed. The Function sets the type of data to be read:

Function

Reads

Disabled

The channel is not used.

1: Read Coil

A 1-bit coil.

2: Read Discrete

A 1-bit discrete input.

3: Read Unsigned Holding (16-bits)

One 16-bit holding register, unsigned.

4: Read Unsigned Input (16-bits)

One 16-bit input register, unsigned.

Read Signed Holding (16-bits)

One 16-bit holding register, signed.

Read Signed Input (16-bits)

One 16-bit input register, signed.

Read Unsigned Holding (32-bits, … register order)

Two holding registers as a 32-bit unsigned value.

Read Unsigned Input (32-bits, … register order)

Two input registers as a 32-bit unsigned value.

Read Signed Holding (32-bits, … register order)

Two holding registers as a 32-bit signed value.

Read Signed Input (32-bits, … register order)

Two input registers as a 32-bit signed value.

Read IEEE-754 Float Holding (32-bits, … register order)

Two holding registers as a 32-bit floating-point value.

Read IEEE-754 Float Input (32-bits, … register order)

Two input registers as a 32-bit floating-point value.

Read IEEE-754 Double Holding (64-bits, … register order)

Four holding registers as a 64-bit floating-point value.

Read IEEE-754 Double Input (64-bits, … register order)

Four input registers as a 64-bit floating-point value.

Each 32-bit and 64-bit function is available in Little Endian and Big Endian register order.

Endianness is the order or sequence of bytes of digital data in computer storage and will be specified by the sensor that is being connected to the UNI-Z1.

A single MODBUS device may have multiple data values that can be read. The Register Address specifies which data the slave device needs to deliver.

In Modbus mode, calibration can be applied so that the registers read by the UNI-Z1 can be scaled to be in the units of what is being measured. For instance, a register that returns 0 to 255 may represent 0% humidity to 100% humidity. The UNI-Z1 can be calibrated to take a number and to convert it to humidity in % and return that as the measured value.

In any system, the sensor and possibly the measured value will be subject to errors that may accumulate to reduce accuracy. In a system that uses the UNI-Z1 to measure fluid volume in a 100 litre tank using a MODBUS sensor, the sensor may have offset errors such that with zero liquid in the tank, the UNI-Z1 is showing a small volume. The UNI-Z1 and sensor may also not be perfectly linear in that they may not measure 1 litre in exactly the same way when the tank is empty versus when it is full. The tank itself may also not be perfectly manufactured and may, for instance have walls that are not perfectly straight. All of these errors could add together such that the final system is less accurate than expected. To achieve a more accurate system, a calibration can be performed. In this example, the tank could be calibrated by adding a small amount of liquid, say 10 litres (low Y) and noting the value reported by the UNI-Z1 (low X). Now fill the tank by adding another 99 litres (high y) and note the value being reported by the UNI-Z1 (high X). By filling the high and low X and Y values into the calibration for that MODBUS channel, offset and non-linearity errors can be eradicated, resulting in a much more accurate system.

In Modbus mode, warning and alarm thresholds can be set for each MODBUS channel. Once enabled, each time a measurement is completed, the returned value will be compared with minimum and maximum warning and alarm thresholds. If a warning or alarm level is breached, a message will immediately be transmitted. As long as the warning or alarm condition persists, messages will be transmitted at the exception-interval rather than the transmit-interval.

Note

If calibration has been applied, then the warning and enable thresholds are in the calibrated units.

A full list of serial interface settings is given in the table at the end of the chapter.

CAN Bus interface

The Senquip UNI-Z1 is equipped with CAN bus that can be used to read data from all kinds of vehicles and sensors. Many sensors, controllers, and electronic devices can be connected to a single CAN network.

In many cases, the protocol that is being used on the CAN bus is known, and so large volumes of understandable data can be extracted from all kinds of vehicles. Common CAN protocols include:

  • J1939, the dominating CAN-based protocol for trucks and busses.

  • ISO 11783, a J1939 flavor for agricultural tractors.

  • ISO 11992, an interface between trucks and trailers.

  • NMEA 2000, a protocol based on J1939 for marine use.

  • CANopen, provides a standard for industrial machinery commonly used in industrial automation.

The CAN interface on the UNI-Z1 supports Classical CAN. Pins 3 and 6 provide the CAN interface, with pin 6 being CAN High (dominant high) and pin 3 being CAN Low (dominant low).

Pin

Function

3

CAN1 L

6

CAN1 H

In CAN networks, 120Ω termination resistors are installed at each end of the network. In most installations, the required termination will already be present and the internal termination resistor should remain disabled. If the UNI-Z1 is located at the end of a CAN network and termination is required, the internal termination resistor can be enabled in the CAN settings. No external termination resistor is required.

Warning

In CAN bus systems, the ground supplied to the Senquip UNI-Z1 must be the same ground as used by the CAN network. High differential voltages between the CAN lines and ground can damage the CAN interface.

Specification

Parameter

Specification

CAN physical layer

High-speed CAN (ISO 11898-2 compatible)

Supported protocols

Classical CAN 2.0 A and 2.0B

Maximum bit rate

1Mbit/s

Receiver common mode range

+/-30V

Bus fault protection on CAN High and CAN Low

+/-58V

ESD protection, CAN bus pins, contact discharge

+/-8kV

ESD protection, CAN bus pins, air discharge

+/-15kV

Optional termination resistor

120Ω

Settings

Measurements can be scheduled as a multiple of the base-interval. The fastest possible measurement rate is achieved by setting the Interval to 1 in which case the CAN network will be sampled on every base interval. To reduce power consumption, the measurement rate can be turned down by increasing the Interval.

The CAN bus peripheral on Senquip devices supports can bit rates of 125kbit/s, 250kbit/s, 500kbit/s and 1Mbit/s as specified in the Nominal Baud Rate field.

Configurable internal 120Ω termination resistors can be enabled.

To ensure minimum intrusion on CAN systems, the CAN peripheral can be set to listen only. In this mode the Senquip device will only receive messages that are acknowledged on the bus by a listening node. Where required, the Senquip device can be made to acknowledge messages by selecting the TX Enable option.

The Capture Time setting can be used to set a timeout after which the CAN bus peripheral will stop listening, allowing the Senquip device to transmit received messages and return to sleep. capture-time can also be used as a mechanism to allow the CAN peripheral to sample the CAN bus for a defined time-period.

A full list of CAN bus settings is given in the table at the end of this chapter.

CAN Filters

Most automotive CAN networks transmit hundreds of different messages, each with a unique identifier. The Senquip device allows you to filter and capture only the messages you need using the ID Capture List.

Basic Usage

Enter the CAN IDs you want to capture in hexadecimal, separated by commas. For example:

18FF20F2,18FF36F0,18FF1BF2

During each measurement interval, the device listens for the listed messages until either all are received or the Capture Time expires.

Capturing Multiple Messages

To capture multiple messages with the same ID, use * followed by the number of each message required:

18FF20F2*4,18FF1BF2*10

This captures:

  • 4 messages with ID 18FF20F2

  • 10 messages with ID 18FF1BF2

To capture all messages of a certain ID received during the interval, use * without a number:

18FF1F12*

Wildcards and Advanced Filters

  • ?? can replace the Priority or Source Address in a J1939 ID.

    For example:

    ??FECA??
    

    This matches any J1939 message with a PGN of FECA, regardless of Priority or Source Address.

  • # can be placed at the end of the list to match one of any other message not already listed:

    18FE1451*4,18FE1256*2,#
    

    This captures:

    • 4 messages with ID 18FE1451

    • 2 messages with ID 18FE1256

    • 1 of every other unique ID not already matched

Special Cases

  • Leave the list blank to capture one of every message that arrives.

  • Use a single * to capture all messages, in the order they arrive.

    Warning

    Capturing all messages may overwhelm the device in high-traffic CAN systems. The number of messages that can be captured in a measurement cycle is internally limited to 128.

Limits

  • Up to 128 messages are captured in each measurement cycle.

  • The ID Capture List can be up to 200 characters long.

Sending CAN Messages

CAN messages can be sent to connected device from within a script. The messages can be sent once, or can be set to repeat at a given time interval. If the device enters sleep or hibernate, repeating messages will be stopped. If the Senquip device is reset, the sending of repeating CAN messages is unaffected. The messages will continue to be sent until a change is applied. For example, if new script is loaded, forcing a reset, the CAN messages will continue to be sent until the new script applies a change. In the case of a malfunction that lasts for more than 30 second, repeating CAN messages will be stopped.

Note

To send CAN messages, the TX Enable option must be selected.

Calibration

Calibration can be applied to each measurement so that the value returned by the Senquip UNI-Z1 is in units that are meaningful in the end application. For instance, a fuel level sensor that outputs a voltage between 0 and 5 V may represent a level between 0 and 100 litres of diesel. The voltage measurement can be calibrated to convert from Volts to the more meaningful unit of litres.

In any system, the measurement instrument (the Senquip UNI-Z1), the sensor and possibly the measured value will be subject to errors that may accumulate to reduce accuracy. In a system that measures fluid volume in a 100 litre tank using a 4-20 mA sensor, offset errors may result in a non-zero or negative reading when the tank is empty. Sensor gain may also not be perfectly linear and so a 1 litre change may be measured differently when the tank is empty versus when it is full. The sensor may report in inches of liquid height where a more meaningful unit may be litres. To achieve an accurate and meaningful measurement, a calibration can be performed.

In this example, the tank could be calibrated by adding a small amount of liquid, say 10 litres at which point the current measured may be 4.1 mA. Now add more liquid to take the level to say 80l. The sensor now reads 16.2 mA. The calibration would then be filled into the IO setting as shown below:

Low In

4.1

This is the value in mA measured by the Senquip UNI-Z1

Low Out

10

This is the actual value that we would like to report

High In

16.2

This is the value in mA measured by the Senquip UNI-Z1

High Out

80

This is the actual value that we would like to report

Unit

l

The unit to be reported is litres

Unit Names and Symbols

The calibration Unit field supports plain text together with a number of commonly used engineering symbols. This allows measurements to be displayed using familiar units such as °C, m³, Ω, µS/cm and N·m.

The following characters are supported:

Letters: a-z, A-Z

Numbers: 0-9

Symbols: /, Ω, ·, µ, Δ, ±, ×, :, °, ², ³, ., %, _, -

Examples:

Unit

Description

V

Volts

A

Amps

°C

Degrees Celsius

°F

Degrees Fahrenheit

Ω

Ohms

kΩ

Kilohms

m³

Cubic metres

m²

Square metres

µS/cm

Microsiemens per centimetre

N·m

Newton metres

kg/m³

Density

%RH

Relative humidity

L/min

Litres per minute

When entering units on Windows, some special characters can be entered using their ALT codes on the numeric keypad.

Character

Windows ALT Code

°

ALT+0176

²

ALT+0178

³

ALT+0179

µ

ALT+0181

±

ALT+0177

×

ALT+0215

The remaining symbols can be copied directly from this guide or inserted using the Windows Character Map (charmap.exe) or the Windows Emoji & Symbols panel(Win+.).

Character

Description

Ω

Ohm

Δ

Uppercase Delta

·

Middle dot

Warnings and Alarms

Each measurement has one alert setting. It is made up of an Alert Type, a Low Threshold, a High Threshold and a Hysteresis. Each time a measurement is completed, it is compared with the thresholds. If it is at or below the low threshold, or at or above the high threshold, an event of the selected type is raised and a message is transmitted immediately. While the event stays active, messages are transmitted at the Exception Interval.

Alert Type

Result when a threshold is crossed

Disabled

No event. The thresholds are ignored.

Info

Event recorded in the device event history; no badge on the Senquip Portal.

Alert

Event shown with a blue indicator.

Warning

Event shown with an orange indicator.

Alarm

Event shown with a red indicator.

The same alert type applies to both thresholds. To check only one threshold, leave the other one blank. For example, for a high-only warning at 50 V, set Alert Type to Warning, leave Low Threshold blank and set High Threshold to 50.

Some alerts, such as the light sensor’s Tamper Alert, have no thresholds: choose the alert type for the event only.

For how each alert type is shown and forwarded by email or SMS, see Exceptions.

Note

If calibration has been applied, set the thresholds in the calibrated units.

Hysteresis

Hysteresis prevents repeated alarms when a measured value fluctuates near a threshold. The hysteresis value is applied on both sides of the configured threshold. For an upper alarm, the value must reach upper threshold + hysteresis to enter the alarm state, then fall to upper threshold - hysteresis to clear it. For a lower alarm, the value must fall to lower threshold - hysteresis to enter the alarm state, then rise to lower threshold + hysteresis to clear it. Between the two switching points, the alarm retains its current state. Hysteresis uses the measurement configured units.

Hysteresis

Upper alarm with a 1.0 V threshold and 0.1 V hysteresis. The alarm starts when the measured value reaches 1.1 V and clears when it falls to 0.9 V. Between those values, the alarm retains its current state. Readings shown in red are in alarm.

External Interface Settings

A full list of external interface settings is given in the table below.

Name

Type

Description

Range or options

Default

Internal Reference

General IO

Overcurrent Protection

On/Off

Enable overcurrent protection on IO terminals. If enabled, the device will disable the IO if the current exceeds a threshold.

On

io.overcurrent_protection

IO1 to IO2

Name

Text

A name for the input that is meaningful to the user.

25 characters

IO1

ioN.name

Interval

Integer

The number of base intervals after which the input is sampled. A value of 1 means that the input is collected every base interval. Set to 0 to disable.

0 to 10000

0

ioN.interval

Wake on Change

On/Off

Wake device from sleep/hibernate on digital change. It is recommended to also enable PULLUP mode for this feature.

Off

ioN.wake

Default State

Option

The state of the output when NOT in an active measurement interval.

OFF, GND, VIN, VSET, PULLUP

OFF

ioN.output.default_state

Measurement State

Option

The state of the output during an active measurement interval.

No Change, GND, VIN, VSET

No Change

ioN.output.meas_state

Measurement Time

Decimal

How long the IO will stay in the Measurement State before measurements are taken.

0 to 3600 Seconds

0

ioN.output.meas_time

Measure Digital Input

On/Off

Enables digital input measurement on terminal.

Off

ioN.digital.enable

Measure Voltage

On/Off

Enables voltage measurement on terminal.

Off

ioN.voltage.enable

Voltage Calibration

Calibration

Calibration parameters. Refer to user guide.

30 characters Volts

0,10,0,10,Volts

ioN.voltage.cal

Voltage Alert

Alert

Alert type, low and high thresholds and hysteresis. See Warnings and Alarms.

ioN.voltage.alert

Measure Current

On/Off

Enables current measurement on terminal.

Off

ioN.current.enable

Optimise for 4-20 mA

On/Off

Use a lower Full Scale Range (FSR) with higher resolution which is best for 4-20 mA devices.

On

ioN.current.low_range

Current Calibration

Calibration

Calibration parameters. Refer to user guide.

30 characters mA

0,10,0,10,mA

ioN.current.cal

Current Alert

Alert

Alert type, low and high thresholds and hysteresis. See Warnings and Alarms.

ioN.current.alert

Measure Frequency

On/Off

Enables frequency measurement on terminal.

Off

ioN.freq.enable

Frequency Calibration

Calibration

Calibration parameters. Refer to user guide.

30 characters Hz

0,10,0,10,Hz

ioN.freq.cal

Frequency Alert

Alert

Alert type, low and high thresholds and hysteresis. See Warnings and Alarms.

ioN.freq.alert

Measure Duty Cycle

On/Off

Enables duty cycle measurement on terminal.

Off

ioN.duty.enable

Duty Cycle Calibration

Calibration

Calibration parameters. Refer to user guide.

30 characters Percent

0,10,0,10,Percent

ioN.duty.cal

Duty Cycle Alert

Alert

Alert type, low and high thresholds and hysteresis. See Warnings and Alarms.

ioN.duty.alert

Measure Pulse Count

On/Off

Enables pulse count measurement on terminal.

Off

ioN.pulse.enable

Pulse Count Calibration

Calibration

Calibration parameters. Refer to user guide.

30 characters Counts

0,10,0,10,Counts

ioN.pulse.cal

Pulse Count Alert

Alert

Alert type, low and high thresholds and hysteresis. See Warnings and Alarms.

ioN.pulse.alert

CAN 1

Name

Text

A name that is meaningful to the user.

25 characters

CAN 1

can1.name

Interval

Integer

The number of base intervals after which the CAN module is turned on. Set to 0 to disable.

0 to 10000

0

can1.interval

Nominal Baud Rate

Integer

Baud rate for CAN communication. Supported values are: 125, 250, 500, 1000

250

can1.nominal_baud

CAN FD Mode

Option

Select CAN FD mode.

Classic, CAN FD, CAN FD BRS

Classic

can1.frametype

Capture Time

Integer

The device will capture matching messages for this length of time.

2

can1.capture_time

TX Enable

On/Off

Allows the device to transmit and acknowledge messages on the CAN bus.

Off

can1.tx_enable

ID Capture List

Text

List of IDs to be captured in HEX format, separated by a comma eg: 18FEE60A. Leave blank to capture all.

200 characters

can1.id_list

Send Raw Data

On/Off

Send all captured data in the standard data message. If unticked, data is only delivered to the scripting handler.

Off

can1.send_data

Termination Enable

On/Off

Enables termination resistor for the CAN bus.

Off

can1.termination

CAN 2

Name

Text

A name that is meaningful to the user.

25 characters

CAN 2

can2.name

Interval

Integer

The number of base intervals after which the CAN module is turned on. Set to 0 to disable.

0 to 10000

0

can2.interval

Nominal Baud Rate

Integer

Baud rate for CAN communication. Supported values are: 125, 250, 500, 1000

250

can2.nominal_baud

CAN FD Mode

Option

Select CAN FD mode.

Classic, CAN FD, CAN FD BRS

Classic

can2.frametype

Capture Time

Integer

The device will capture matching messages for this length of time.

2

can2.capture_time

TX Enable

On/Off

Allows the device to transmit and acknowledge messages on the CAN bus.

Off

can2.tx_enable

ID Capture List

Text

List of IDs to be captured in HEX format, separated by a comma eg: 18FEE60A. Leave blank to capture all.

200 characters

can2.id_list

Send Raw Data

On/Off

Send all captured data in the standard data message. If unticked, data is only delivered to the scripting handler.

Off

can2.send_data

Serial 1

Name

Text

A name for the input that is meaningful to the user.

25 characters

Serial 1

serial1.name

Interval

Integer

The number of base intervals after which the serial port is turned on. Set to 0 to disable.

0 to 10000

0

serial1.interval

Type

Option

The electrical interface type.

RS232, RS485

RS232

serial1.type

Termination Resistor

On/Off

This parameter enables the integrated termination resistor.

Off

serial1.termination

Mode

Option

Describes how the serial port is to be handled. CAPTURE: serial data is captured between start and end characters. MODBUS: serial data is treated according to MODBUS RTU standard

Capture, Modbus, Scripted

Capture

serial1.mode

Baud Rate

Integer

Baud rate for serial communication. Common values are: 4800, 9600, 19200, 38400, 57600, 115200

57600

serial1.baud

Settings

Text

A string describing the number of bytes: 7,8,9. Parity type: N(none), E(even), O(odd). Number of stop bits: 1 or 2. Typically: 8N1

8N1

serial1.settings

Capture

Start String

Text

The serial port starts reading data when it detects these characters. Example: $GPGGA, serial data will be ignored until $GPGGA is received after which data will be captured. If nothing is specified, the serial port will capture all data until the timeout period is reached.

32 characters

serial1.capture.start

Idle Time Before Start

Integer

For a valid start condition, there must be this amount of idle time before receiving serial data. Additionally, the captured data will restarted if the serial port is idle for this time. Set to 0 to disable.

0 to 60000 Milliseconds

0

serial1.capture.start_idle_time

End String

Text

Once capturing, if these characters are received, the serial port will stop capturing and will return to sleep. For binary data or escape sequences refer to the User Guide.

32 characters

serial1.capture.end

Request String

Text

This string will be sent when the serial port is first turned on. Use this function to request data from a remote module.

32 characters

serial1.capture.request

Max Time

Integer

The device will wait this length of time for a valid capture.

5

serial1.capture.maxtime

Max Chars

Integer

Maximum number of characters to be captured before the serial port goes back to sleep.

to 2000

500

serial1.capture.maxchars

Capture Alert

Alert

Alert type for this event. See Warnings and Alarms.

serial1.capture.alert

Modbus RTU

Slave Timeout

Decimal

How long to wait for a response from each slave device.

0.1 to 10 Seconds

0.2

serial1.modbus.timeout

Delay Between Reads

Integer

Delay between successive read commands.

10 to 10000 Milliseconds

15

serial1.modbus.framedelay